skip to content

UV-Vis Spectrometer

Technical Data

UV-Vis Spectrometer

Company: Perkin Elmer LAS GmbH
Model: Lambda 35
Measuring principle: Wavelength-dependent light absorption and excitation spectra (PL)
Sample preparation: Solid samples: Size mind. 10 mm & thickness max. 5 mm
Liquids: Standard cuvettes
Type of construction: Double beam, split beam, transmission, reflection
Radiation source: Halogen and deuterium lamp
Wavelength range: 190…1100 nm
Spectral bandwidth: 0,5,1,2,4 nm
Reproducibility: ± 0.05 nm
Wavelength accuracy: ± 0.05
Scattered light (%T): At 220 nm ≤ 0.01%
At 340 nm ≤ 0.01%